MEDAID GLOBAL RECOMMENDATIONS FOR INSTRUMENTS CLEANING

About Stainless Steel:
Surgical instruments manufacturing involve almost 70% stainless steel as a basic raw material. The term “stainless steel“ is subject to water spotting and staining. The most common problem is a stain cause by a surface deposit, which appears in a number of colors. But once the cause is found and eliminated , the problem diminishes completely.
Stainless steel used for surgical instrument manufacturing involve a number of reasons like it is rust resistant, it can be honed to an extremely sharp edge or fine point. Also it can be hardened for delicacy and preciseness.
In order to ensure stainless steel corrosion resistance, a special two-step corrosion process is adopted. But only these procedures alone cannot guarantee the most useful possible life of the instruments.

Water used for Decontamination and Cleaning:
The drinking water may contain high concentrations of chemical elements that can cause the deterioration of instruments during cleaning. In particular high chloride concentration are to be considered dangerous because they lead to the forming of corrosion holes. The chloride content must be kept as low as possible e.g., not more than 100 mg/l, otherwise corrosion risks become very high.

Corrosion and Stains:
Corrosion results from the action of a liquid or gas reagent on a metal or alloy, therefore, risks of corrosion require preventive action regular tests throughout the life of the instruments.

Water Stains and Halos:
Halos are colored surface marks without accurately defined contours. They result from the presence of mineral substances or heavy metal ions in water or sterilization steams. Water stains are similar phenomena but, in general, their contours are sharp. They result from the residue of organic substances in water or from an excessive concentration of mineral substances such as limestone.
Deposits can be eliminated by energetic friction or, if necessary, using a non-abrasive cleaner. The best method is to eliminate the mineral salt from the washing or sterilization water (or to use distilled or de-mineralized water for flushing and pure water steam for sterilization). If the deposits are not eliminated , pitting corrosion may occur.
There may also be whitish stains of phosphates, which is characteristic of some cleaning products.

Color Surface Residue:
Some yellowish brown to dark brown residue is sometimes wrongfully considered to be rust stains appearing on sterilized instruments. Essentially, this is to be found in the least accessible areas during cleaning. Again, energetic rubbing or the use of a cleaning product will remove the stains without leaving any marks. These stains are due to residue, which had set in solidly and resisted cleaning and sterilization.
Instruments preserved in disinfecting solutions not renewed often can also be coated with deposits. During sterilization, these deposits become scale and take on a dark brown color.

Oxidization:
Oxidization is unusual on surgical instruments except if they are soaked for long periods of time in concentrated acid solutions. In this case, the surface of the instrument takes on a rusted look: this comes from the formation of an iron oxide film in places. Such oxidization may rapidly degenerate and become pitting corrosion.

Pitting Corrosion:
This is the most frequent form of corrosion and consists in the forming of small but often, deep holes in material whose surface is not deeply attached elsewhere. Such corrosion is caused above all by halogen ions (chlorides, iodine, bromides) working on the surface and resulting from the saline solutions or other chlorides, from iodine or from dirty disinfecting solutions. Even if contact with halogens is difficult to avoid, to prevent this type of corrosion, instruments should be cleaned as soon as they have been used. Pitting corrosion is often associated with multi-colored stains resulting from organic residue or other incorrectly eliminated residue or that is associated with isolated plaques of corrosion.

Localized Corrosion Plaques:
The mechanism causing such corrosion is similar to that of pitting corrosion and shares many of its characteristics. It appears more particularly on surfaces at articulation points of forceps.

Corrosion by the Forming Crevasses or Caving:
This is often the case of metal parts that come into contact with other parts of the same metal or of another non-metallic material, an assembly liable to form an occluded zone: this is often developed in the hidden parts of the instruments. Essentially it appears by the outcrop of rust appearing in narrow cracks. However, it is often easily confused with the residue of operations or cleaning products. Precise identification is only possible by laboratory analysis.

Corrosion by Friction or Inter-Granular Corrosion:
Austenitic and ferritic stainless steel may be affected by corrosion more particularly at grain joints. The matrix remains unaffected and they often cause the total breakdown of the metal without such a phenomenon being suspected. Therefore, it occurs between the interstices of two parts moving in opposite directions (such as scissors). The cause is a lack of lubrication: dry friction of two parts against one another causes microscopic chips to form, and rust sets in.

Corrosion by Contact:
The corrosion comes from the colse contact between two metals of different compositions in the presence of electrolyte, for instance water. At the contact points, rust stains appear. The corrosion can become very extensive if stainless steel instruments are mixed with regular steel instruments with incomplete or worn out surface protection (e.g., old chrome-plated or nickel-plated instruments whose chrome or nickel film has come away).

Corrosion by Fracturing Under Stress:
In the presence of stresses of various origins (welding mechanical or heat treatment) and in chlorinated media, austentic steels undergo serve transgranular fracturing; ferric steels are less sensitive to this type of corrosion. To avoid damage,  it is indispensable keep the instruments in an open position throughout the cleaning process and for instruments with hooks or racks, not to go beyond the first notch during sterilization. These stresses produced by the temperature build-up followed by cooling, may provoke clefting or deterioration at the articulation points. However, it is necessary to differentiate between this clefting and the fracturing by the forcing of the instrument.

Surface Corrosion:
Surface corrosion is particularly rare but may be caused by the contact between instruments and strong acids or caustic solutions or by the deposit of surrounding or suspended rust.

Suspended Rust:
In some of the aforementioned cases, rust form in cracks, clefts or at contact surfaces. If these instruments are not eliminated systematically, degradation will increase on each new utilization cycle. During washing, the rust formed in this way will be transmitted to the other instruments by cleaning products or by sterilization products (rust in suspension) and cause increasing surface corrosion on the later.

Various Rust Phenomena:
If the steam production installations or piping are in poor conditions, suspended rust particles may appear in the steam. After sterilization, corrosive precipitation may appear at the surface of the instruments or on the conditioning.

COLORED HALO STAINS WITH SHARP CONTOURS
Problem Deposits or organic substances
Possible Cause
Solution Rinse instrument throughly and immediately after use in distilled or de-mineralized water.

COLORED HALO STAINS AND THE SHARP CONTOURS ARE NOT PRESENT
Problem Metal deposits
Possible Cause In flushing water or steam there are iron, copper, manganese etc are present.
Solution: Use distilled or de-mineralized water.

WHITISH STAINS
Problem Phosphate deposits.
Possible Cause
Solution Check composition of cleaning solutions.

YELLOW TO DARK BROWN COLORS
Problem Residue Deposits
Possible Cause Old cleaning solutions / incorrectly cleaned instruments.
Solution Change baths after each use / rinse instruments immediately after use in distilled or de-mineralized water.

BLACK STAINS
Problem Caustic reaction
Possible Cause Detergent with low pH content (6 or less).
Solution Use neutral detergent.

BLUISH-BLACK STAINS
Problem Metallic coating or painting.
Possible Cause Metals of different types have been sterilized together such as, for instance stainless steel with silver / chrome / ...
Solution Sterilize only the similar metal instruments together.

SEVERAL COLORS
Problem Excess heat (Chromatic Oxide).
Possible Cause The autoclave is not operating properly or has been set to very high temperature (this can cause metal to lose some of its properties such as strength and hardness, etc.)
Solution Correct autoclave temperature.
How  to Develop a Comprehensive Instrument Care Program:
Proper instrument care can increase the useful life of the instrument possible. Proper cleaning, lubrication, correct handling and storage can increase the useful life of the instrument. In order to ensure the longest useful life possible, regularly inspect your instrument for sign of corrosion, stain and water.

Storage and Cleaning:
Instruments must be left in original packing if not use immediately.
Rough handling should avoid and make sure to avoid contact as it may cause scratching of the surface and may damage to finish.
Sudden or hard impacts may cause heat-treated instruments with cutting edges to crack.
Instruments must be completely dry.
Sharp edges should not touch each other.
Protect the tip of sharps.
Face all jaws the same way. Set cupped instruments so water cannot collect.
Unlike metal instruments should keep separate.

Cleanliness:
Cleanliness is the prerequisite in the care of surgical instruments. Any type of material left on the surface of the instruments can lead to corrosion. Blood, chlorides, and slat leave residue that can clog working mechanism. Make sure the instrument is dry throughly to prevent moisture. Lubricate the instruments before autoclave and rinse with distilled water. During using the instrument always avoid contact with cutting edges.

Ultrasonic Cleaning:
Ultrasonic cleaning is the most effective process to clean the instruments. For mixing of the solution, manufacturer recommendation should be followed. While instrument cleaning, do not mix different metal instruments in the same cycle. Make sure that instruments having rachets and box locks are accessible.
If there is any stain or deposits occur remove these with a quality polish. But make sure that you do not use abrasive powders or steel wool.

Sterilizing Process:
Instruments made up of metals can be sterilized through steam, flash, dry heat or gas. To avoid the problems during sterilizing process, make sure that there is no rust of foreign material in steam pipes. Also the door seal must be very tight. Using distilled water should carry sterilizing process.

Proper Handling:
Proper handling can increase the useful life of the instrument when it is used. Make sure there should not be any instrument abuse. MedAid Global strongly recommend that the instrument should be used only for designed purpose. Forceps like bone cutting, and ronguers should only be used on bones.
Pin or wire cutter should be used on metal.
Scissors are the particularly subject to abuse. Misuse can easily damage the cutting edge and blades as well.
Eye scissors are very delicate and should be used only for the purpose it is made.
Pliers and needle holders are two different instruments and should not be used in place of one to other. For needle holders, always match the needle holder to the size of the needle being used.
Haemostatic forceps should not be used in place of needle holders. As the haemostats are used as a blood vessel clamps only.

MEDAID GLOBAL CLEANING RECOMMENDATIONS FOR THE INSTRUMENTS:
After a research of many years and experience of many years, MedAid Global has developed the following recommendations for cleaning of its instruments. We ensure you that if you will follow our policy for cleaning, instruments will have a longer useful life.

1. Soaking
After the instruments have been used, dip the instruments in a quality soak. But before putting, make sure about blood, tissue or other debris from hardening in serration’s, box joints and moving parts.

2. Brush the Instruments
After step one the instruments must be on hand cleaned in a solution using nylon brush. While brushing the instruments, make sure that your attention is to the box locks, serrations and other areas of the instruments.

3. Cleaning
For proper cleaning the instruments use only quality detergents or liquid recommended for this particular purpose. Make sure the instrument is in open position when you clean it.

4. Inspection
After the cleaning process, the instrument must be throughly inspected. Make sure that all the moving parts of the instruments are moving smoothly. Also inspect sharp edges.

5. Lubrication
The purpose of lubrication is to prevent the moving parts from freezing so that your instruments are in good working condition. The lubricant you use should be good water-soluble.

6. Sterilization
After cleaning and lubrication the instruments must be dried before sterilization. Autoclave is best recommended for sterilization method.